Fluke 125 Fluke 125 Users Manual - Page 98
Dual Input Meter, Advanced Scope Functions, Input A and Input B
View all Fluke 125 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 98 highlights
Fluke 125 Users Manual Advanced Scope Functions Display Modes Normal ....... Captures up to 40 ns glitches and displays analog-like persistence waveform. Smooth ...... Suppresses noise from a waveform. Envelope.... Records and displays the minimum and maximum of waveforms over time. Auto Set (Connect-and-View™) Continuous fully automatic adjustment of amplitude, time base, trigger levels, trigger gap, and hold-off. Manual override by user adjustment of amplitude, time base, or trigger level. Dual Input Meter The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1%. More than one waveform period must be visible on the sceen. Input A and Input B DC Voltage (VDC) Ranges 500 mV, 5V, 50V, 500V, 1250V Accuracy 0.5% +5 counts) Normal Mode Rejection (SMR 60 dB @ 50 or 60 Hz ±1% Common Mode Rejection (CMRR)........ >100 dB @ DC >60 dB @ 50, 60, or 400 Hz Full Scale Reading 5000 counts True RMS Voltages (VAC and VAC+DC) Ranges 500 mV, 5V, 50V, 500V, 1250V Accuracy for 5 to 100% of range DC coupled: DC to 60 Hz (VAC+DC 1% +10 counts) 1 Hz to 60 Hz (VAC 1% +10 counts) 10-4