Fluke 125 Fluke 125 Users Manual - Page 41
Isolated Triggering, TRIG:A, FREE RUN, ON TRIG., Isolated Triggering
View all Fluke 125 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 41 highlights
j Accept all trigger selections and return to normal measurement. Note Setting the automatic triggering to >1Hz will slow down the auto range. TRIG:A appears in gray text on bottom of the screen when no trigger is found. Note Gray text in a menu or button bar indicates that the function is disabled or the status is not valid. FREE RUN: the test tool automatically updates the trace even if there are no triggers. ON TRIG.: the screen is updated only when valid triggers occur. 2 Scope/Meter Mode Triggering on a Waveform Isolated Triggering Use the optically isolated trigger probe (ITP120, optional) to trigger on an external source, and to isolate the test tool from a trigger waveform. See Figure 2-16. To choose the isolated trigger probe, select 'EXT' in point e of the previous example. Trigger level is fixed and is TTL compatible. Figure 2-16. Isolated Triggering 2-19