Fluke 199C/S FE 192,196,199 C Users Manual - Page 101
Troubleshooting, The Test Tool Shuts Down After A Few Seconds
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 101 highlights
Troubleshooting The Test Tool Does Not Start Up • The batteries may be completely empty. In this case the test tool will not start up, even if it is powered by the battery charger. Charge the batteries first: power the test tool with the battery charger without turning it on. Wait about 15 minutes and try turning on the test tool again. The Test Tool Shuts Down After A Few Seconds • The batteries may be empty. Check the battery symbol at the top right of the screen. A symbol indicates that the batteries are empty and must be charged. The Screen Remains Black • Make sure that the test tool is on. • You might have a problem with the screen contrast. Press , then press . Now you can use the arrow keys to adjust the contrast. 8 Maintaining the Test Tool Troubleshooting The Operation Time Of Fully Charged Batteries Is Too Short • The batteries may be in poor condition. Refresh the batteries to optimize the condition of the batteries again. It is advised to refresh the batteries about four times a year. The Printer Does Not Print • Make sure that the interface cable is properly connected between the test tool and the printer. • Make sure that you have selected the correct printer type. (See Chapter 6.) • Make sure that the baud rate matches with the printer. If not, select another baud rate. (See Chapter 6.) • If you are using the PAC91 (Print Adapter Cable), make sure that it is turned on. 93