Fluke 199C/S FE 192,196,199 C Users Manual - Page 71
Finding Missing Pulses, Trigger Options, Trigger on Pulse Width, Condition, >t, Update, On Trigger
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 71 highlights
Finding Missing Pulses The next example covers finding missing pulses in a train of positive pulses. In this example it is assumed that the pulses have a 100 ms distance between the rising edges. If the time accidently increases to 200 ms, a pulse is missing. To set the test tool to trigger on such missing pulses, let it trigger on gaps bigger than about 150 ms. Do the following: 1 Display the TRIGGER key labels. 2 Open the Trigger Options menu. 5 Triggering on Waveforms Triggering on Pulses 4 Select the positive pulse icon to trigger on the gap between the positive pulses, then jump to Condition. 5 Select >t, then jump to Update. 6 Select On Trigger. The test tool is now prepared to trigger on pulse gaps. Observe that the trigger menu at the bottom of the screen has been adapted to set the pulse condition: 3 Select Pulse Width on A... to open the Trigger on Pulse Width menu. 63