Fluke 199C/S FE 192,196,199 C Users Manual - Page 111

Diode, Temperature TEMP, DC Voltage VDC, AC+DC Voltage True RMS, DMM Measurements on Meter Inputs

Page 111 highlights

Diode Maximum Voltage Reading 2.8 V Open Circuit Voltage 4 V Accuracy 2 % +5 counts) Measurement Current 0.5 mA, ±20 % Temperature (TEMP) With Optional Temperature Probe Ranges (°C or °F 40.0 to +100.0 ° -100.0 to +250.0 ° -100.0 to +500.0 ° -100 to +1000 ° -100 to + 2500 ° Probe Sensitivity 1 mV/°C and 1 mV/°F DC Voltage (VDC) Ranges ..... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V Full Scale Reading 5000 counts Accuracy 0.5 % +5 counts) Normal Mode AC Rejection at 50 or 60 Hz ±1 % .... >60 dB 9 Specifications DMM Measurements on Meter Inputs AC Voltage (VAC) Ranges......500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V Full Scale Reading 5000 counts Accuracy 15 Hz to 60 Hz 1 % +10 counts) 60 Hz to 1 kHz 2.5 % +15 counts) For higher frequencies the frequency roll off of the Meter input starts affecting accuracy. Normal Mode DC Rejection 50 dB AC+DC Voltage (True RMS) Ranges......500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V Full Scale Reading 5000 counts Accuracy DC to 60 Hz 1 % +10 counts) 60 Hz to 1 kHz 2.5 % +15 counts) For higher frequencies the frequency roll off of the Meter input starts affecting accuracy. All accuracies are valid if the waveform amplitude is larger than 5 % of full scale. 103

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122

Specifications
DMM Measurements on Meter Inputs
9
103
Diode
Maximum Voltage Reading
......................................
2.8 V
Open Circuit Voltage
.................................................
<4 V
Accuracy
...............................................
±
(2 % +5 counts)
Measurement Current
...............................
0.5 mA,
±
20 %
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°
C or
°
F)
..................................
-40.0 to +100.0
°
-100.0 to +250.0
°
-100.0 to +500.0
°
-100 to +1000
°
-100 to + 2500
°
Probe Sensitivity
............................
1 mV/
°
C and 1 mV/
°
F
DC Voltage (VDC)
Ranges
.....
500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading
........................................
5000 counts
Accuracy
............................................
±
(0.5 % +5 counts)
Normal Mode AC Rejection at 50 or 60 Hz
±
1 %
....
>60 dB
AC Voltage (VAC)
Ranges
......
500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading
........................................
5000 counts
Accuracy
15 Hz to 60 Hz
..................................
±
(1 % +10 counts)
60 Hz to 1 kHz
...............................
±
(2.5 % +15 counts)
For higher frequencies the frequency roll off of the
Meter input starts affecting accuracy.
Normal Mode DC Rejection
...................................
>50 dB
AC+DC Voltage (True RMS)
Ranges
......
500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading
........................................
5000 counts
Accuracy
DC to 60 Hz
......................................
±
(1 % +10 counts)
60 Hz to 1 kHz
...............................
±
(2.5 % +15 counts)
For higher frequencies the frequency roll off of the Meter
input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is
larger than 5 % of full scale.