Fluke 199C/S FE 192,196,199 C Users Manual - Page 110
Meter, DMM Measurements on Meter Inputs, Meter Input, Meter Functions, General, Continuity CONT
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 110 highlights
Fluke 192B - 196B/C - 199B/C Users Manual Meter Meter Input Input Coupling DC Frequency Response DC to 10 kHz (-3 dB) Input Impedance 1 MΩ (±1 %)//10 pF (±1.5 pF) Max. Input Voltage 1000 V CAT II 600 V CAT III (For detailed specifications, see "Safety") Meter Functions Ranging Auto, Manual Modes Normal, Relative 102 DMM Measurements on Meter Inputs The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. General DC Common Mode Rejection (CMRR 100 dB AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB Ohms (Ω) Ranges 500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ Full Scale Reading 500 Ω to 5 M 5000 counts 30 M 3000 counts Accuracy 0.6 % +5 counts) Measurement Current 0.5 mA to 50 nA, ±20 % decreases with increasing ranges Open Circuit Voltage 4 V Continuity (CONT) Beep 50 Ω (±30 Ω) Measurement Current 0.5 mA, ±20 % Detection of shorts of 1 ms