Fluke 199C/S FE 192,196,199 C Users Manual - Page 27
Variable Input Sensitivity, Variable
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
Variable Input Sensitivity The variable input sensitivity allows you to adjust the input A sensitivity continuously, for example to set the amplitude of a reference signal to exactly 6 divisions. The input sensitivity of a range can be increased up to 2.5 times, for example between 10 mV/div and 4 mV/div in the 10 mV/div range. To use the variable input sensitivity, do the following: 1 Apply the input signal 2 Perform an Auto Set (AUTO must appear at the top of the screen) An Auto Set will turn off the variable input sensitivity. You can now select the required input range. Keep in mind that the sensitivity will increase when you start adjusting the variable sensitivity (the displayed trace amplitude will increase). 3 Display the INPUT A key labels. 1 Using The Scope Acquiring Waveforms 4 Open the Input A Options... menu. 5 Select and accept Variable. 6 Exit the menu. At the bottom left of the screen the text A Var is displayed. Selecting Variable will turn off cursors and automatic input ranging. 7 Press mV to increase the sensitivity, press V to decrease the sensitivity. 19