Documenting Process Calibrator Specifications DC Voltage Output Range Resolution % of Output + % of Full Scale 1 Year 2 Year 110 mV 1 µV 0.01% + 0.005% 0.015% + 0.005% 1.1 V 10 µV 0.01% + 0.005% 0.015% + 0.005% 15 V 100 µV 0.01% + 0.005% 0.015% + 0.005% Temperature Coefficient: (0.001% of output + 0.001% of f.s.)/°C in the ranges -10 to 18 °C and 28 to 50 °C Maximum Output Current: 10 mA Loading: (0.001% f.s. + 1 µV)/ mA Common Mode Error: 0.008% f.s/(Common Mode Volt) Maximum Input Voltage: 30 V dc 109
Documenting Process Calibrator
Specifications
109
DC Voltage Output
Range
Resolution
% of Output + % of Full Scale
1 Year
2 Year
110 mV
1
μ
V
0.01% + 0.005%
0.015% + 0.005%
1.1 V
10
μ
V
0.01% + 0.005%
0.015% + 0.005%
15 V
100
μ
V
0.01% + 0.005%
0.015% + 0.005%
Temperature Coefficient:
(0.001% of output + 0.001% of f.s.)/
°
C in the ranges -10 to 18
°
C and 28 to 50
°
C
Maximum Output Current:
10 mA
Loading:
(0.001% f.s. + 1
μ
V)/ mA
Common Mode Error
: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage:
30 V dc