Fluke 744 Fluke 744 Users Manual - Page 82
damping. The error of the expected measured, value is shown in the top left of the measure
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 82 highlights
744 Users Manual 16. Press the Done softkey to accept the calibration parameters. The display changes to: gj46s.eps 17. You now have the choice of starting an automatic test or stepping through the test points manually. Press the Auto Test softkey to have the calibrator run through the tests automatically. (Abort gets you out of the calibration procedure.) The tests begin at the zero point, sourcing the correct temperature (a voltage) and measuring the corresponding current from the transmitter. As soon as a measurement has settled and been captured, the calibrator moves to the next step. Because the calibrator waits for the 72 measurement to stop changing, the Auto Test works correctly for instruments with built-in damping. The error of the expected measured value is shown in the top left of the measure window. 18. The calibrator moves to the remaining set of points. For temperature and electrical parameter calibration, the points are done automatically. If you are sourcing pressure, the calibrator pauses at each step for you to adjust the pressure source. When the tests are complete, an error summary table such as the following is displayed. gj47s.eps