To avoid damaging the pressure module, from overpressure, never apply pressure
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 41 highlights
Gage 700P06 PRESSURE MODULE RANGE 100 PSIG 700 kPa 7 bar BURST PRESSURE 300 PSIG Low Differential 700P04 PRESSURE MODULE RANGE 15 PSID/G 100 kPa 1 bar BURST PRESSURE 45 PSIG High gj11f.eps Figure 11. Gage and Differential Pressure Modules Documenting Process Calibrator Using Measure Mode Caution To avoid mechanically damaging the pressure module, never apply more than 10 ft.-lb. of torque between the pressure module fittings, or between the fittings and the body of the module. Always apply appropriate torque between the pressure module fitting and connecting fittings or adapters. To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the pressure module. To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the printing on the pressure module or the pressure module instruction sheet for the acceptable material compatibility. 31