Measurement Scale, Linear-Output Transmitters
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 51 highlights
Measurement Scale This feature lets you scale the measurements in accordance with a particular process instrument's response. Percent of scale works for linear-output transmitters or square-law transmitters such as differential pressure transmitters that report flow rate. Linear-Output Transmitters 1. If necessary, press M for MEASURE mode. 2. Select a measurement function (m,v, h,q, t, or p) as previously described. 3. Press the Scale softkey. 4. Select % scale from the list. 5. Use the numeric keypad to enter the 0% of scale value (0% Value). 6. Press e. 7. Use the numeric keypad to enter the 100% of scale value (100% Value). 8. Press e. Documenting Process Calibrator Using Measure Mode 9. Press the Done softkey. Percent of scale remains in effect until you change to another measurement function or press the Scale softkey and select another scale mode. 41