Fluke 744 Fluke 744 Users Manual - Page 90

The calibrator is now in Transmitter mode. It is, measuring the process input and sourcing

Page 90 highlights

744 Users Manual 7. Press M for SOURCE mode. 8. Press the appropriate function key for the control output (e.g., v or m). If the transmitter is connected to a current loop that has a power supply, select Simulate Transmitter for the current output choice. 9. Select a source value, e.g., 4 mA. 10. Press M for MEASURE/SOURCE mode. 11. Press More Choices until the Transmitter Mode softkey appears. 12. Press the Transmitter Mode softkey. 13. Set the 0% and 100% values for MEASURE and SOURCE on the display. You can select Linear or √ for the transfer function. 14. Press Done. 15. The calibrator is now in Transmitter mode. It is measuring the process input and sourcing the control signal output proportional to the input. 16. To change any of the Transmitter mode parameters, press Change Setup, and repeat the process in step 13. 17. To exit Transmitter mode, press the Abort softkey. 80

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138

744
Users Manual
80
7.
Press
M
for SOURCE mode.
8.
Press the appropriate function key for the control
output (e.g.,
v
or
m
). If the transmitter is
connected to a current loop that has a power
supply, select
Simulate Transmitter
for the
current output choice.
9.
Select a source value, e.g., 4 mA.
10. Press
M
for MEASURE/SOURCE mode.
11. Press
More Choices
until the
Transmitter Mode
softkey appears.
12. Press the
Transmitter Mode
softkey.
13. Set the 0% and 100% values for MEASURE and
SOURCE on the display. You can select
Linear
or
for the transfer function.
14. Press
Done
.
15. The calibrator is now in Transmitter mode. It is
measuring the process input and sourcing the
control signal output proportional to the input.
16. To change any of the Transmitter mode
parameters, press
Change Setup
, and repeat the
process in step 13.
17. To exit Transmitter mode, press the
Abort
softkey.