Table 1. Summary of Source and Measure Functions, Function, Measure, Source, Short, m - 120
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
744 Users Manual Table 1. Summary of Source and Measure Functions Function Measure Source vdc V 0 V to +/-300 V 0 V to 15 V (10 mA max) hac V 0 V to 300 V rms, 20 Hz to 5 kHz No sourcing hFrequency 1 Hz to 1 kHz (100 mV to 300 V rms) 0.1 V to 10 V p-p sine wave, or peak square 1 kHz to 30 kHz (0.5 V to 30 V rms) wave, 0 Hz to 50 kHz 30 kHz to 50 kHz (1 V to 30 V rms) qResistance 0 Ω to 11 k Ω 0 Ω to 11 k Ω mdc Current 0 mA to 110 mA 0 to 22 mA (28 V max), sourcing or sinking qContinuity Beep and the word Short indicates continuity No sourcing tThermocouple Types E, N, J, K, T, B, R, S, C, L, or U tRTD 100 Ω Platinum (3926) 100 Ω Platinum (385) 120 Ω Nickel (672) 200 Ω Platinum (385) 500 Ω Platinum (385) 1000 Ω Platinum (385) 10 Ω Copper (427) 100 Ω Platinum (3916) pPressure 27 modules ranging from 0 to 10 in. H O 2 (2.5 kPa) to 0 to 10,000 psi (69,000 kPa) Note sLoop Power 24 or 28 V (22 mA max) Note: Use an external hand pump or other pressure source as a pressure stimulus for the source pressure function. 4