Fluke 744 Fluke 744 Users Manual - Page 125
Fluke 744 Manual
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 125 highlights
Documenting Process Calibrator Specifications Temperature, Thermocouples (cont) Type Range °C Measure °C Source °C 1 Year 2 Year 1 Year 2 Year U -200 to 0 0.6 0.9 0 to 600 0.3 0.4 0.4 0.6 0.3 0.4 Sensor inaccuracies not included. Accuracy with external cold junction; for internal junction add 0.2°C Resolution: 0.1°C Temperature Scale: ITS-90 or IPTS-68, selectable Compensation: ITS-90 per NIST Monograph 175 for B,R,S,E,J,K,N,T; IPTS-68 per IEC 584-1 for B,R,S,E,J,K,T; IPTS-68 per DIN 43710 for L,U. Temperature Coefficient: 0.05°C/°C in the range -10 to 18°C and 28 to 50°C Common Mode Error: 0.01°C/(Common Mode Volt) Maximum Input Voltage: 30 V 115
Documenting Process Calibrator
Specifications
115
Temperature, Thermocouples (cont)
Type
Range
°
C
Measure °C
Source °C
1 Year
2 Year
1 Year
2 Year
U
-200 to 0
0.6
0.9
0.4
0.6
0 to 600
0.3
0.4
0.3
0.4
Sensor inaccuracies not included.
Accuracy with external cold junction; for internal junction add 0.2
°
C
Resolution:
0.1
°
C
Temperature
Scale:
ITS-90 or IPTS-68, selectable
Compensation:
ITS-90 per NIST Monograph 175 for B,R,S,E,J,K,N,T; IPTS-68 per IEC 584-1 for B,R,S,E,J,K,T;
IPTS-68 per DIN 43710 for L,U.
Temperature Coefficient
: 0.05
°
C/
°
C in the range -10 to 18
°
C and 28 to 50
°
C
Common Mode Error:
0.01
°
C/(Common Mode Volt)
Maximum Input Voltage:
30 V