Fluke 744 Fluke 744 Users Manual - Page 45

Table 5. Thermocouple Types Accepted, Positive Lead, Color, Negative Lead, Specified Range, Material

Page 45 highlights

Documenting Process Calibrator Using Measure Mode Table 5. Thermocouple Types Accepted Type Positive Lead Material Positive Lead (H) Color ANSI* IEC** Negative Lead Material E Chromel Purple Violet Constantan N Ni-Cr-Si Orange Pink Ni-Si-Mg J Iron White Black Constantan K Chromel Yellow Green Alumel T Copper Blue Brown Constantan B Platinum (30% Rhodium) Gray Platinum (6% Rhodium) R Platinum (13% Rhodium) Black Orange Platinum S Platinum (10% Rhodium) Black Orange Platinum C *** Tungsten (5% Rhenium) White Tungsten (26% Rhenium) L (DIN J) Iron Constantan U (DIN T) Copper Constantan *American National Standards Institute (ANSI) device negative lead (L) is always red. **International Electrotechnical Commission (IEC) device negative lead (L) is always white. *** Not an ANSI designation but a Hoskins Engineering Company designation. Specified Range (°C) -250 to 1000 -200 to 1300 -210 to 1200 -270 to 1372 -250 to 400 600 to 1820 -20 to 1767 -20 to 1767 0 to 2316 -200 to 900 -200 to 600 35

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138

Documenting Process Calibrator
Using Measure Mode
35
Table 5. Thermocouple Types Accepted
Type
Positive Lead
Positive Lead (H)
Color
Negative Lead
Specified Range
Material
ANSI*
IEC**
Material
(
°
C)
E
Chromel
Purple
Violet
Constantan
-250 to 1000
N
Ni-Cr-Si
Orange
Pink
Ni-Si-Mg
-200 to 1300
J
Iron
White
Black
Constantan
-210 to 1200
K
Chromel
Yellow
Green
Alumel
-270 to 1372
T
Copper
Blue
Brown
Constantan
-250 to 400
B
Platinum (30% Rhodium)
Gray
Platinum (6% Rhodium)
600 to 1820
R
Platinum (13% Rhodium)
Black
Orange
Platinum
-20 to 1767
S
Platinum (10% Rhodium)
Black
Orange
Platinum
-20 to 1767
C ***
Tungsten (5% Rhenium)
White
Tungsten (26% Rhenium)
0 to 2316
L (DIN J)
Iron
Constantan
-200 to 900
U (DIN T)
Copper
Constantan
-200 to 600
*American National Standards Institute (ANSI) device negative lead (L) is always red.
**International Electrotechnical Commission (IEC) device negative lead (L) is always white.