Fluke 744 Fluke 744 Users Manual - Page 45
Table 5. Thermocouple Types Accepted, Positive Lead, Color, Negative Lead, Specified Range, Material
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 45 highlights
Documenting Process Calibrator Using Measure Mode Table 5. Thermocouple Types Accepted Type Positive Lead Material Positive Lead (H) Color ANSI* IEC** Negative Lead Material E Chromel Purple Violet Constantan N Ni-Cr-Si Orange Pink Ni-Si-Mg J Iron White Black Constantan K Chromel Yellow Green Alumel T Copper Blue Brown Constantan B Platinum (30% Rhodium) Gray Platinum (6% Rhodium) R Platinum (13% Rhodium) Black Orange Platinum S Platinum (10% Rhodium) Black Orange Platinum C *** Tungsten (5% Rhenium) White Tungsten (26% Rhenium) L (DIN J) Iron Constantan U (DIN T) Copper Constantan *American National Standards Institute (ANSI) device negative lead (L) is always red. **International Electrotechnical Commission (IEC) device negative lead (L) is always white. *** Not an ANSI designation but a Hoskins Engineering Company designation. Specified Range (°C) -250 to 1000 -200 to 1300 -210 to 1200 -270 to 1372 -250 to 400 600 to 1820 -20 to 1767 -20 to 1767 0 to 2316 -200 to 900 -200 to 600 35