Fluke 744 Fluke 744 Users Manual - Page 42
or Pa. Metric, O, bar, g/cm - calibration procedure
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 42 highlights
744 Users Manual 1. Connect a pressure module to the calibrator as shown in Figure 12. The threads on the pressure modules accept standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary. 2. Press M for MEASURE mode. 3. Press p. The calibrator automatically senses which pressure module is attached and sets its range accordingly. 4. Zero the pressure module as described in the module's Instruction Sheet. Modules vary in zeroing procedures depending on module type. You MUST perform this step before you execute a task that sources or measures pressure. 5. If desired, you can change pressure display units to psi, mHg, inHg, mH2O, inH2O@, inH2O@60°F, ftH2O, bar, g/cm2, or Pa. Metric units (kPa, mmHg, etc.) are shown in Setup mode in their base units (Pa, mHg, etc.). Change the pressure display units as follows: a. Press s. b. Press Next Page twice. c. Press eor the Choices softkey with the cursor on Pressure Units. d. Select the pressure units with u or d. e. Press e. f. Press Done. 32