Fluke 744 Fluke 744 Users Manual - Page 66
Simulating RTDs, Temperature Detector
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 66 highlights
744 Users Manual Simulating RTDs Note Refer to Table 6 for information about RTD (Resistance-Temperature Detector) types supported by the calibrator. Connect the calibrator to the instrument under test as shown in Figure 21. The figure shows connections for two, three, or four-wire transmitters. For three or four-wire transmitters, use the 4-inch long stackable jumper cables to connect the third and fourth wires at the source V Ω RTD jacks. Proceed as follows to simulate an RTD (ResistanceTemperature Detector): 1. If necessary, press M for SOURCE mode. 2. Press t until the select RTD type display is showing. 3. Press the u or d keys followed by e to select the desired RTD type. 4. Enter the temperature you want to simulate as prompted by the display, then press e. 56