Fluke 744 Fluke 744 Users Manual - Page 56

Using Source Mode, Sourcing Electrical Parameters - test leads

Page 56 highlights

744 Users Manual Using Source Mode The operating mode (i.e., MEASURE, SOURCE) is shown in a reverse-video bar on the display. If the calibrator is not in SOURCE mode, press M until SOURCE is shown. You must be in SOURCE mode to change any of the SOURCE parameters. Sourcing Electrical Parameters To select an electrical sourcing function, proceed as follows: 1. Connect the test leads as shown in Figure 16, depending on the source function. 2. Press M for current, v for dc voltage, h for frequency, or q for resistance. 3. Enter the desired output value, then press e. For example, to source 5.0 V dc, press v 5 . 0 e. 46 Note If you are sourcing frequency, respond to the display prompt to select a zero-symmetric sine or positive square wave. The amplitude you specify is p-p amplitude. 4. To change the output value enter a new value and press e. Note If you are sourcing current, wait for the g symbol to go out before you use the output. 5. To set the output value to 0 in the present source function, press c. 6. To turn off sourcing completely, press c twice. Note Use the source current function to drive a current loop. This is different than the loop power function in which the calibrator is powering a process instrument. To source loop power, use the Loop Power function accessible from Setup mode.

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138

744
Users Manual
46
Using Source Mode
The operating mode (i.e., MEASURE, SOURCE) is
shown in a reverse-video bar on the display. If the
calibrator is not in SOURCE mode, press
M
until
SOURCE is shown. You must be in SOURCE mode
to change any of the SOURCE parameters.
Sourcing Electrical Parameters
To select an electrical sourcing function, proceed as
follows:
1.
Connect the test leads as shown in Figure 16,
depending on the source function.
2.
Press
M
for current,
v
for dc voltage,
h
for frequency, or
q
for resistance.
3.
Enter the desired output value, then press
e
. For example, to source 5.0 V dc, press
v
5
.
0
e
.
Note
If you are sourcing frequency, respond to the
display prompt to select a zero-symmetric
sine or positive square wave. The amplitude
you specify is p-p amplitude.
4.
To change the output value enter a new value
and press
e
.
Note
If you are sourcing current, wait for the
g
symbol to go out before you use the output.
5.
To set the output value to 0 in the present source
function, press
c
.
6.
To turn off sourcing completely, press
c
twice.
Note
Use the source current function to drive a
current loop. This is different than the loop
power function in which the calibrator is
powering a process instrument. To source
loop power, use the
Loop Power
function
accessible from Setup mode.