Fluke 744 Fluke 744 Users Manual - Page 48
orientation. See Attach the RTD to input jacks as the display or
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 48 highlights
744 Users Manual To measure temperature using an RTD input, proceed as follows: 1. If necessary, press M for MEASURE mode. 2. Press t twice so that Select RTD Type shows. 3. Press u or d to select the desired RTD type. 4. Press e. 5. Press u or d to select a 2-, 3-, or 4- wire connection. 6. Attach the RTD to input jacks as the display or Figure 15 shows. Use the supplied jumper between the mA Ω RTD MEAS low jack and the V MEAS low jack as shown if you are using a 3wire connection. 7. Press e. Caution Do not force a dual banana plug between any two jacks in the horizontal orientation. Doing so will damage the jacks. Use the supplied jumper wire when needed for RTD measurements. You can use a dual banana plug in the vertical orientation. See Figure 14. 38