Fluke 744 Fluke 744 Users Manual - Page 122
Frequency Sourcing, Range, Accuracy, and 2 Year, Waveform Choices, Amplitude, Amplitude Accuracy
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 122 highlights
744 Users Manual Frequency Sourcing Range Accuracy 1 and 2 Year 0.00 Hz to 10.99 Hz 11.00 Hz to 109.99 Hz 0.01 Hz 0.01 Hz 110.0 Hz to 1099.9 Hz 1.100 kHz to 21.999 kHz 22.000 kHz to 50.000 kHz 0.01 Hz 0.002 kHz 0.005 kHz Waveform Choices: Zero-symmetric sine or positive square wave, 50% duty cycle. Amplitude: 0.1 to 10 V pk Amplitude Accuracy: 0 Hz to 1099 Hz: 3% of output + 0.5% f.s 1.1 kHz to 10.9 kHz: 10% of output + 0.5% f.s 11 kHz to 50 kHz: 30% of output + 0.5% f.s Maximum Input Voltage: 30 V dc 112
744
Users Manual
112
Frequency Sourcing
Range
Accuracy
1 and 2 Year
0.00 Hz to 10.99 Hz
0.01 Hz
11.00 Hz to 109.99 Hz
0.01 Hz
110.0 Hz to 1099.9 Hz
0.01 Hz
1.100 kHz to 21.999 kHz
0.002 kHz
22.000 kHz to 50.000 kHz
0.005 kHz
Waveform Choices
: Zero-symmetric sine or positive square wave, 50% duty cycle.
Amplitude
: 0.1 to 10 V pk
Amplitude Accuracy
:
0 Hz to 1099 Hz
: 3% of output + 0.5% f.s
1.1 kHz to 10.9 kHz
: 10% of output + 0.5% f.s
11 kHz to 50 kHz
: 30% of output + 0.5% f.s
Maximum Input Voltage: