Fluke 744 Fluke 744 Users Manual - Page 84
“As Left” Test Run, as left
View all Fluke 744 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 84 highlights
744 Users Manual 4. Adjust the transmitter output for 20 mA. 5. If the span was adjusted in step 4, you must go back and repeat steps 3 and 4 until no more adjustment is required. 6. Now check the transmitter at 50%. If it is within specification, your adjustment is complete. If not, adjust the linearity and begin this procedure again at step 3. "As Left" Test Run Proceed as follows to generate and record as left data for the thermocouple temperature transmitter you have just adjusted. 1. Press the As Left softkey to record as left data. 2. Press the Auto Test softkey to begin an automatic sequence through all the test points, or you can step through the tests manually. 3. When the tests are complete, observe the error summary table, such as the following. gj48s.eps An asterisk (*) next to a measure or source value indicates an unsettled value (g annunciator) when the measurement was taken. 4. If all the results are within specification, as they are this time, press the Done softkey. An entry in memory is made for as left data. 74