Fluke 744 Fluke 744 Users Manual - Page 83

Adjusting the Transmitter, adjustment controls and connection points for your

Page 83 highlights

19. In the results summary test, failures are highlighted. An adjustment is required in this example because three tests show failures. The failures were outside the ±0.5% tolerance that we selected. 20. Either press the Done softkey to save the data, or the Abort softkey to delete the data and start over. You can see the saved data entry and recall the table for later viewing through the Review Memory softkey during normal operation. You can upload this data to a host computer running compatible application software. Documenting Process Calibrator Calibrating a Process Instrument Adjusting the Transmitter Proceed as follows to make the calibration adjustments to the transmitter. (Always refer to the transmitter manufacturer's instructions to locate the adjustment controls and connection points for your transmitter.) 1. Press the Done softkey while viewing the results summary. 2. Press the Adjust softkey. The calibrator sources 0% of span (100°C in this example) and displays the following softkeys: • Go to 100%/Go to 0% • Go to 50% • As Left • Exit Cal 3. Adjust the transmitter output for 4 mA then press the Go to 100% softkey. 73

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138

Documenting Process Calibrator
Calibrating a Process Instrument
73
19. In the results summary test, failures are
highlighted. An adjustment is required in this
example because three tests show failures. The
failures were outside the ±0.5% tolerance that
we selected.
20. Either press the
Done
softkey to save the data,
or the
Abort
softkey to delete the data and start
over.
You can see the saved data entry and recall the
table for later viewing through the
Review
Memory
softkey during normal operation. You
can upload this data to a host computer running
compatible application software.
Adjusting the Transmitter
Proceed as follows to make the calibration
adjustments to the transmitter. (Always refer to the
transmitter manufacturer’s instructions to locate the
adjustment controls and connection points for your
transmitter.)
1.
Press the
Done
softkey while viewing the results
summary.
2.
Press the
Adjust
softkey. The calibrator sources
0% of span (100°C in this example) and displays
the following softkeys:
Go to 100%/Go to 0%
Go to 50%
As Left
Exit Cal
3.
Adjust the transmitter output for 4 mA then press
the Go to 100% softkey.